Operating Instructions Chapter 2
DME4000
8014584/ZN33/2017-07 © SICK AG • Germany • Subject to change without notice 73
Operation
Information field 1 (example)
65 dB
Reception level: current reception signal
Ver. 1.0
Software Version
3.3.3.1
Menu subitem (see the menu overview); parameter changes not possible
3.3.3.1 !
“!”: Enter code correctly: Parameter changes possible
3.3.3.1 ?
“?”: Entered parameter is invalid; previously valid parameter remains stored
Cause: for example, lower switching limit > upper switching limit
Correction measure: correct parameters
Information field 2 (example)
+ 010.000 m
Current measurement value; is displayed during measurement operation.
Code
Menu items and parameters are displayed here. “Code” appears
when you press .
Service
Blinking Service display signals a hardware defect or
too high or too low temperature
Keys
Prompts through the menu structure, code, parameter entry, storing of
parameters
Exit a menu subitem
Selection of menu items and entry of numbers