Operating manual PA
31
AI.V2 = 1200 (temperature 1200 °C)
7.3 General functions (configuration layer C011)
Green Status LED
You can assign specific functions to the LED:
LED is continuously lit to indicate 24 V operating voltage
LED indicates status of switching output 1
LED indicates status of switching output 2
LED indicates ATD function in progress
Set the function using parameter LED.6.
Activate Laser
If your PA pyrometer features a built-in laser, there are several ways in
which it can be activated. Configure parameter PILO with the particu-
lar method you have selected:
Activate using button on rear panel
Permanently off
Edge-triggered: based on signal input (0 -> 24V transition)
Use parameter PIL.T to configure the laser ON-time between 1 (mini-
mum) and 15 minutes (maximum).
NOTE !
Do not attempt to operate the laser continuously. The built-in laser
is not designed for continuous use. When the user-defined laser
auto-shutoff time is reached (maximum programmable duration =
15 minutes) the laser will automatically switch off. When the am-
bient temperature exceeds 55°C the laser will automatically deac-
tivate.
Video camera settings
Pyrometer models with an integrated video camera feature the following
operating modes.
Target Brightness Control (TBC) settings