48
13.1 Calibration/scaling via CellaView
To use the user-defined calibration function, activate it first in expert
mode.
Start CellaView
Open the menu Settings Extras -> Settings
Select expert mode and activate editable calibration
Close the menu
Open the menu Pyrometer settings
The parameters can now be edited under the tab Spectral channel 1.
13.2 Calibration/scaling via terminal connection
If ever required, the ARDOCELL PA can be recalibrated using the sub-
menu Calibration. Press command „K“ and then enter the password
„100“ to access the calibration menu.
The calibration menu opens.
------------------------------------------------------------------
------------------------------------------------------------------
Submenu CALIBRATION
------------------------------------------------------------------
Name .... "Pyrometer PA Series"
1: [LAMBDA 1 CALIBRATION]
A: Reset settings to factory default
S: Set pyrometer name
Z: End Calibration-Mode
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>CALIBRATION >
------------------------------------------------------------------
Submenu LAMBDA 1
------------------------------------------------------------------
L1 range .... 0.0 - 1000.0 C
L1 User calibration ....... off
L1 User def. offset +0.00000
L1 User def. factor +1.00000
A: Set L1 - extended-range
B: Set L1 User-Cal. On/Off
C: Set L1 User-Cal. Offset
D: Set L1 User-Cal. Factor
ESC: Back to MAIN-MENU
------------------------------------------------------------------
>CALIBRATION >LAMBDA 1 >
You can reset all configurations you have made to your ARDOCELL PA
and restore the factory default settings using Command „A“. This also