Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic
02.05 CS SD 24
General information 13
Page 13 of 48
Medical Solutions
Parts conducting voltage after system OFF and/or with power
plug pulled out
0
Main system control board D1 and power board D2 0
WARNING
Danger high voltage!
When the system is switched off, there is still electrical voltage on
boards D1 and D2 of the main system.
Disregarding safety precautions can result in death or serious
bodily injury.
¹ Therefore before working on one of these boards, make
sure that the LED V400 has gone out. This is the case 3
minutes after the ARCADIS Varic system is switched off.
Monitor cart, monitors and image system 0
WARNING
Danger high voltage!
When the system is switched off and even when the power plug is
pulled out, the line voltage is still applied to the power assembly of
the monitor cart, to the monitors and to the image system.
Disregarding safety precautions can result in death or serious
bodily injury.
¹ Before work is done on the monitor cart, the power plugs
must be pulled out of the UPS.
WARNING
Dangerous radiation during checks and adjustments!
Risk of death or serious bodily injury!
¹ For checks and adjustments that must be performed with
the radiation switched on, the prescribed radiation safety
measures must be observed; if necessary, wear radiation
protective clothing (see also ARTD-002.731.02.xx.xx and
ARTD-002.731.38.xx.xx “General Guidelines for Techni-
cal Service”).
X ¹ These checks and adjustments are explicitly designated
on the following pages with the radiation warning sym-
bol.