8 General information
ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG
02.05 CS SD 24
Page 8 of 48
Medical Solutions
Notes on adapting the power plug to local conditions. 0
The customer can have the power plug on the supplied power cable replaced with an
appropriate local plug by an electrician, provided that:
The power plug used can conduct the power required for the operation of the ARCADIS
Varic (refer to current and voltage values stated on the line voltage label, back of moni-
tor cart).
The required line internal resistance is reached (refer to ARCADIS Varic Planning
Guide).
The selectivity of the fuse protection is assured - according to the relevant national
standards.
It must also be ensured by the customer that connection to an unsuitable power supply is
prevented (e.g. by identification or design measures).
As concluding measures, perform and document protective ground wire measurements
(in Germany e.g. in the medical device book).

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals