18 General Start-up
ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG
02.05 CS SD 24
Page 18 of 48
Medical Solutions
Function check of the emergency stop switches 0
Connect the system to the voltage and boot it.
Click "Emergency" for the patient registration.
Operate the EMERGENCY STOP switch on the C-arm.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
Disengage the EMERGENCY STOP switch.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop (C-arm movement and emergency stop / p. 40).

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors