30 Specific system settings
ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG
02.05 CS SD 24
Page 30 of 48
Medical Solutions
Make country-specific changes and click <Save>.
Click <Home>.
If a change is made to the fluoro time, check this function.
Country-specific adaptation of the maximum tube current 0
NOTE
A maximum tube current limit is required - as far as known - only
in Denmark.
Start service program and log in as described under "Service Login".
Fluoroscopy alarm Fluoroscopy blocking
USA
Service mask
input in 1/10 min.
4.5 min.
45
5 min.
5
Europe
Service mask
input in 1/10 min.
5 min.
50
10 min.
50