Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic
02.05 CS SD 24
Specific system settings 29
Page 29 of 48
Medical Solutions
Fig. 13: Main System_Configuration_Buzzer
• Under Configuration click <Buzzer>.
Make changes according to the following description:
• Block time settings
- Warning time limit: The buzzer sound starts when the entered radiation time has
elapsed.
- Block time limit: After the entered radiation time has elapsed - without reset of the
“warning time limit" - the radiation is blocked the next time it is selected. Every reset
of the “warning time limit” also resets the “block time limit” to “0”.
• Buzzer mode (direct, cassette)
- After exposure:Buzzer sound after the exposure.
- During exposure:Buzzer sound during the exposure.
• Buzzer mode (indirect, I.I.)
- Buzzer off:Buzzer sound off. Exceptions: Push mode and “warning time limit” has
ended.
- Buzzer on: Buzzer sound always during radiation.
The country-specific regulations must be observed for the max. fluoroscopy time and the
radiation blockage.
Factory setting: