Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic
02.05 CS SD 24
General Start-up 21
Page 21 of 48
Medical Solutions
Testing the ground wire 0
The system must be free of voltage.
Measuring de-
vice:
Ground wire testing device; e.g. safety tester Unimed 1100
Measuring de-
vice:
Max. 0.2 Ohm (observe country-specific regulations!) (current:
10 A)
Test procedure: Measure between all touchable, conductive parts of the system
and the ground wire at the ARCADIS Orbic power plug.