Siemens SP00-000.814.09.01.02 ARCADIS
11.07 CS PS SP
Installing the WLAN Expansion 23
Page 23 of 78
Medical Solutions
• Press the power switch into the break-out. See (Fig. 14 / p. 22).
¹ Installation location of the switch: 0 label = top; 1 label = bottom.
Wiring of the WLAN Power Supply Power Cable - Power Switch
• Lay the cable from the WLAN power supply through the round opening
(4/Fig. 13 / p. 21) up to the inside of the top compartment to the WLAN power switch.
• Connect the cable with the designation M15 L on the WLAN power supply, terminal L
(5/Fig. 13 / p. 21).
• Connect the cable with the designation M15 N on the WLAN power supply, terminal N
(5/Fig. 13 / p. 21).
• Plug in the cable with the designation S15.1A on the power switch (1/Fig. 14 / p. 22).
• Plug in the cable with the designation S15.2A on the power switch (2/Fig. 14 / p. 22).
NOTE
The loose ground wire covered with heat-shrink tubing that
is from the power cable does not have to be connected at
both ends.
Wiring the WLAN Power Switch with the Power Outlet Strip
• Lay the cable from the power outlet strip up to the WLAN power switch.
• Plug in the cable with the designation S15.1 on the power switch (3/Fig. 14 / p. 22).
• Plug in the cable with the designation S15.2 on the power switch (4/Fig. 14 / p. 22).
• Plug in the three-prong connector at the power outlet strip.
NOTE
The loose ground wire from the power cable (switch side)
does not have to be connected. A ground wire connection to
the monitor carriage is already connected.
Wiring the Network Cable
• Plug in the new network cable on the RJ45 socket on the WLAN Client
(9/Fig. 13 / p. 21) and lay it to the side to the existing network switch.
• Plug in the other end of the network cable at the next free socket on the network switch.
• Secure the cable at suitable locations using cable ties.
Labels
• Attach one of the labels with the warning symbol for non-ionizing radiation and a refer-
ence symbol to to the Operator Manual along side each other next to the WLAN power
switch.
• Attach the 2nd label with the warning symbol for non-ionizing radiation and the 2nd ref-
erence symbol to the Operator Manual on the top console of the monitor carriage,
directly below the cable outlet for the mouse (= location of the signal source below the
console, centered to the signal antennas of the WLAN Client).