24 WLAN Configuration
DRAFT 22.04.2008
ARCADIS SP00-000.814.09.01.02 Siemens
11.07 CS PS SP
Page 24 of 78
Medical Solutions
3- 3WLAN Configuration
WLAN Primary Setup 0
NOTE
The network connection between the WLAN Client and the
image system PC can be established using the Primary
Setup Tool. For this, the Primary Setup Tool must be
installed from the CD on the image system PC.
Other programming is then performed using the integrated
HTML server of the WLAN Client.
Installation of the Primary Setup Tool 0
Logging off as the "Meduser" and Logging in as the "Administrator"
NOTE
Installation of the Primary Setup Tool requires administrator
privileges on the image system PC.
Fig. 16: Utilities shutdown application
The system is booted up and and ready.
Start local service and enter the password.
Click on <Utilities> in the Main menu.
Select "System" under <Utilities> - <System>.
Select "Shutdown Application" and click on <Go>.
¹ The message “Application Shutdown was selected. Are you sure? appears.