44 WLAN Configuration
DRAFT 22.04.2008
ARCADIS SP00-000.814.09.01.02 Siemens
11.07 CS PS SP
Page 44 of 78
Medical Solutions
NOTE
If the IP address / Subnet mask for the WLAN Client needs to
be changed during operation, this can be done by using the
Primary Setup Tool.
The "DLC Protocol" and "PROFINET IO RTA Protocol"
installed during installation of the PST Tool must also be
activated, because they are deactivated again following
assignment of the WLAN Client IP address and Subnet mask.
Deactivating Network Protocols no longer needed 0
NOTE
The "DLC Protocol" and "PROFINET IO RT Protocol" net-
work protocols installed to perform the PST Tool do not
need to be active for further operation.
If the IP address / Subnet mask for the WLAN Client needs to
be changed during operation, this can be done by using the
Primary Setup Tool. The deactivated "DLC Protocol" and
"PROFINET IO RT Protocol" network protocols first must be
activated again temporarily.