EasyManuals Logo

Siemens Maxum II Reference Manual

Siemens Maxum II
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
2.2 Power Entry and Control Module
2.2.1 PECM Overview
Overview
The PECM3-CTL board mounts on the PECM-SSR board. This assembly provides a variety
of power and control functions. The connections are shown below.
AC in
Optional
UPS input
for 24vdc
supply
LWH1 - LWH5
out
MWH out
Purge Air
Switch
24v in
Fan
power
HWH SSR
control out
I
2
C bus
I
2
C bus
Solenoid
control out
Atmospheric
pressure sensor
Air bath heater
monitor in
(Ribbon-cable
connector to
PECM-SSR)
Purge signal out
LWH/MWH control in
HWH control out
HWH control in
HW
H
co
nt
cont
r
ol
HW
HW
HW
H
H
H
co
co
co
nt
nt
nt
nt
nt
ro
ro
ro
ro
l
contr
o
TL/OT modules
Temp RTD in
Filtered
AC out
to 24v
supply
L1 MMI
LED out
MWH
HWH SSR
power out
PECM-SSR Board
PECM3-CTRL Board
F1
F1
F5
F2
F5
F5
F2
F2
F2
F2
F2
F2
F2
F2
F2
F2
F2
F2
F2
F4 (without cover)
B
oa
rd
cov
er)
(without cov
F4 (witho
F4 (without
F3 (without cover)
Purge disable jumper
Figure 2-4 PECM3 I/O Connections
The PECM3 assembly part number is 2021828-002. An upgrade kit, part number 2022019-001
is available to replace earlier units.
Electronic Compartment Component Descriptions and Maintenance Procedures
2.2 Power Entry and Control Module
Maxum II Reference Manual
18 Manual, 7/2017, 2000596-001

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals