2.2.3.6 Electronic Enclosure Environment
Purge Monitoring
The PECM monitors the state of the purge condition for the analyzer. If a loss of purge is
detected the purge switch is enabled. The purge control alarm signal is controlled by the
SYSCON. The purge signal cable from SYSCON to PECM plugs into connector J1302 on the
PECM2. Connection SW1 on the PECM2 is used to connect atmospheric reference for the
purge switch.
When a purged enclosure is not required per the safety codes, connector J2 on the PECM2
can be used to disable the purge alarm. See the
PECM3 I/O Connections
diagram in
PECM
Overiew
for connector locations.
Atmospheric Pressure Monitoring (New for PECM-CTL3)
This sensor allows a Maxbasic program to measure the ambient atmospheric pressure for
custom applications. A tube must be connected from the sensor (J44 on the PECM-SSR board)
to the exterior of the EC.
L1 MMI LEDs
Maintenance Panel Level 1 consists of LEDs on the outside of the analyzer door. It is intended
for use in GCs that are not equipped with the full feature Maintenance Panel. The PECM
supplies the control signals for Maintenance Panel Level 1, if equipped. For PECM-1, the
Maintenance Panel Level 1 connects to position J17. See the
PECM3 I/O Connections
diagram in
PECM Overiew
for the location of connector J17.
Electronic Compartment Component Descriptions and Maintenance Procedures
2.2 Power Entry and Control Module
Maxum II Reference Manual
26 Manual, 7/2017, 2000596-001

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals