Configuration
The SNE is configured by the System Controller (SYSCON) and periodically reports analysis
results. It can be interactively controlled for Real-Time decisions on operation scenarios. The
SNE software controls all sampling relating to its internal configuration and sensor setup.
Components
The major SNE software components are as follows:
Data Manager
Communications Manager
Hardware Manager
Computational Engine
Data Manager
The Data Manager maintains configuration data that controls hardware sequence of events
and controls what manipulation is performed on sampled data.
The Data Manager also provides results and status information to externally connected devices
via the Communication Manager. This data is organized as a set of Sensor Analyzer Module
(SAM) structures. The data represents the unit as a standard sensor to external host.
Communication Manager
The Communication Manager acts as a central point of control for communication links
attached to the Sensor Near Electronics (SNE). This allows Internal SNE software to function
regardless of which communication link is being used to communicate with the system.
Hardware Manager
The Hardware Manager provides scheduling and communication services for the hardware in
the analysis zone. These include devices such as the following:
Detectors
Sample valves
Relays
Pressure monitors and controllers
Temperature monitors and controllers
Flow control valves
Computational Engine
The Computational Engine takes acquired chromatography data and performs system
calculations. Most of these calculations are performed by functions contained in the EZChrom
method, which provides all peak identifications and integration and response factors.
Electronic Compartment Component Descriptions and Maintenance Procedures
2.6 Sensor Near Electronics (SNE) Software
Maxum II Reference Manual
Manual, 7/2017, 2000596-001 73

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals