Display messages 2 - 15
Siemens-Elema AB SPR8-215.898.01 Page 15 of 22 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
CAL 28 Generator ON
during start-up
An error was
detected during pro-
gram execution. The
program is inter-
rupted.
• Check that all supply voltages on
D1 are present, refer to page 110**.
• If CAL 28 occurs again or remains
displayed: the ROM containing the
program may be defective.
Replace the software (IC’s I72, I66
and I74) on D1 ; refer to page
110**.
• If the error remains after replacing
the software, D1 should be
replaced.
• Enter the cause and the error
message "CAL 28" in the service
protocol.
CAL 29 Generator ON The single tank is
not correctly con-
nected.
• The single tank is missing or the
cable harness is incorrectly
connected.
• Check that connection K9 on D1 is
connected.
• Check for possible damage in the
main cable harness between D1
and the single tank. Measure K9:
13-4 and move the arm and single
tank while the connections are
being measured. See page 140**.
If damage is detected, replace the
main cable harness (both cables).
Refer to the section "Replacement
of the arm cable harness".
Dis-
played
messages
Mode of
operation
Cause Possible reason
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E