Mechanical systems 10 - 25
Siemens-Elema AB SPR8-215.898.01 Page 25 of 28 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Replacing the vertical friction linings 10
Raise the single tank to maximum height and secure it with a ceiling support (Fig. 28).
Fig. 28
Never work with the system open if the capacitor is charging or
has just been charged.
The capacitor bank may still be charged even if the system is
switched OFF and the mains cable is disconnected.
Life-threatening electric shock hazard exists.
See Chap.:
"Protective measures when working on the capacitor bank".
Equipment required: Large caliper, rope, ceiling support.
The ceiling suspension must be able to hold a weight of 100 kg,
since the entire weight of the arm will be on the ceiling support
once the spring is removed.
If not, the arm and single tank will fall.
Make sure the ceiling support is sufficient.
Do not stand under the single tank or the arm system during the
work that follows.
MOB00484
Min. 2400
DANGER
CAUTION
NOTE

Table of Contents

Other manuals for Siemens MOBILETT Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus and is the answer not in the manual?

Siemens MOBILETT Plus Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals