EasyManua.ls Logo

Siemens MOBILETT Plus - List of Displayed Messages

Siemens MOBILETT Plus
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Display messages 2 - 21
Siemens-Elema AB SPR8-215.898.01 Page 21 of 22 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
List of displayed messages 2
* Max. preparation time is 10 s if the FW version is lower than 1.2 or 1.2E.
Err 3 Overvoltage >142 kV
Err 4 Poor kV regulation
Err 5 No tube current
Err 6 Anode speed too slow
CAL 7 Exposure time too long
___ ___
Err 9 High-tension asymmetry
USE 10 Single tank >55° or <15°C, measured at the location of the temperature sensor
___ ___
CAL 12 Values in CMOS memory lost
CAL 13 Filament voltage, 60 V DC on D4 not present
CAL 14 Overvoltage in the capacitor bank (>360 V)
USE 15 Preparation level held by the user >20 s*
Err 16 Capacitor bank charging too fast
Err 17 Capacitor bank charging too slow
Err 18 Capacitor voltage charging too slowly
CAL 19 Leakage current too high in the capacitor
USE 20 Radiation release button was released during the exposure
Err 21 Filament current incorrect for stand-by
Err 22 Filament current incorrect for exposure
Err 23 During exposure, the main inverter was interrupted >3 times
USE 24 The selected kV and mAs values are lost
USE 25 The calculated anode load is lost
CAL 26 Error in RAM
CAL 27 Error in ROM
CAL 28 Error in the operating system
CAL 29 No single tank connected or an error in the connection
USE 30 The customer values stored in CMOS memory are lost
CAL 31 Replace backup battery <2.8 V.
___ ___
___ ___
Err 34 Error log overfull
Err 35 Watchdog responded

Table of Contents

Other manuals for Siemens MOBILETT Plus

Related product manuals