Service programs 5 - 3
Siemens-Elema AB SPR8-215.898.01 Page 3 of 10 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
If an error is detected, the test program will be aborted, an audible beep will be signaled
and a message will be displayed:
If the test ends successfully, the display shows:
P02 - Rotation test, if the FW version is lower than 1.5 or 1.5E 5
(Wiring diagram X037E, page 150)
Accelerate the anode with a 150 Hz current during two seconds. After this first accelera-
tion, the speed is measured and if necessary, a second acceleration of two seconds
occurs. When the correct speed is reached, the anode will rotate freely for 10 seconds*
before it is braked during 7 seconds** to a stop. An error message will be displayed after
two unsuccessful attempts to accelerate the anode. When the test is activated, the display
shows:
After acceleration, the display shows the actual rotation frequency: If an error is detected,
the test program will be aborted, an audible beep will be signaled and a message will be
displayed:
If the test ends successfully, the display shows:
* 20 seconds if FW lower than V1.2 or V1.2E
** 9 seconds if FW lower than V1.4 or V1.4E
kV mAs
P02 Err 6 = Rotation error
kV mAs
P02 PASS = Rotation test passed in program P02
kV mAs
Rot tESt = Rotation test
kV mAs
Rot 143 = Rotation speed 143 Hz: Normal value 149 Hz. Min. value 142 Hz.
kV mAs
P02 Err 6 = Rotation error
kV mAs
P02 PASS = Rotation test passed in program P02