EasyManua.ls Logo

Siemens MOBILETT Plus - Collimator Replacement and Repairs; Collimator Lamp Replacement

Siemens MOBILETT Plus
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 - 1
Siemens-Elema AB SPR8-215.898.01 Page 1 of 8 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Replacing the collimator lamp 9
Remove the cover (1) by removing the screws (4) marked "A" (Fig. 1). Turn the collimator
around so that the lamp housing is in front of the cover opening.
Remove the protective plate (2), covering the collimator lamp by loosening, not
removing, the screws marked "B" one turn. Do not adjust screws "C", "D", or "E" (Fig. 2).
Remove the defective lamp and replace it with a new lamp (Fig. 3). Do not touch the
bulb with bare fingers. (See "list of spare parts".)
Reinstall the protective plate. Tighten the screws marked "B".
Fig. 1 Fig. 2
Fig. 3
Equipment required: Lux meter
75
75
75
75
90
90
7
5
75
7
5
75
6
0
60
4
5
45
3
0
30
15
15
0
1
5
15
30
30
45
45
6
0
60
7
5
75
9
0
90
7
5
75
MOB00475
B
D
E
B
2
C
MOB00474
A
A
A
A
1
Replacement and repairs on the collimator 9

Table of Contents

Other manuals for Siemens MOBILETT Plus