MOBILETT Plus / - E/ - M SPR8-215.898.01 Page 4 of 8 Siemens-Elema AB
Rev. 05 01.02 SPS-UD Solna, Sweden
9 - 4 Replacement and repairs on the collimator
Replacing the collimator 9
• Switch off the unit and disconnect the power plug.
• Turn the single tank so that the collimator is facing upwards (Fig. 7).
• Secure the arm and the single tank with a rope (Fig. 7).
• Remove covers "A" and "B".
• Remove collimator cover "C".
Fig. 7
Equipment required: Rope
C
A
B