Replacement and repairs on the collimator 9 - 3
Siemens-Elema AB SPR8-215.898.01 Page 3 of 8 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Set a SID of 100 cm (39"). Use a measuring tape.
Switch on the light localizer and align the cassette. Collimate the light field to
25 x 25 cm (10" x 10").
Mark the four corners with coins and mark one of the corners with two coins (Fig. 5).
Set 52 kV, 2.0 mAs on the control panel.
Release an exposure.
Develop the film and measure the deviation between the light and the radiation fields.
Acceptable deviation ± 1.7% of SID, (Fig. 6).
If the deviation is more than 1.7%, adjust with the screws marked C and D (Fig. 2), to
move the light field closer to the radiation field.
Repeat the test with a new exposure.
If necessary, adjust again until the deviation between the light field and the radiation field
is within the tolerance specified.
Reattach cover 1 (Fig. 1).
X-ray radiation!
If not observed, risk for exposure to radiation.
Use radiation protection!