General 1 - 3
Siemens-Elema AB SPR8-215.898.01 Page 3 of 6 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Safety information and protective measures 1
• When performing service work and tests adhere to:
- the product-specific safety information in the documents,
- as well as the general safety information contained in ARTD
Part 2.
• Connect the MOBILETT Plus/Plus E/Plus M only to mains power
supply outlet (receptacle) , that corresponds to the installation
requirements of VDE 0107 or country-specific regulations.
• Remove or install boards only when the generator is switched
OFF. Adhere to the ESD guidelines .
• Checks and adjustments performed with radiation ON are
identified by the radiation warning symbol . During these
types of adjustments, radiation protective clothing must be
worn.
• Never work with the system open if the capacitor is charging or
has just charged.
• The capacitor bank may still be charged even if the system is
switched OFF and the mains cable is disconnected. Life-
threatening electric shock hazard exists.
To avoid danger, refer to the section entitled
“Protective measures when working on the capacitor bank”.