Replacement and repairs on the collimator 9 - 5
Siemens-Elema AB SPR8-215.898.01 Page 5 of 8 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Disconnect the cable to the connection terminal board "D" (Fig. 8).
If there are cable ties, cut them (Fig. 8).
Remove the collimator by loosening the four screws "E" (Fig. 8). Reserve the hardware.
Install the new collimator. Center the rotary-flange so that the screws marked "E" are
centered in the grooves before tightening.
Connect the wires to connection terminal board "D".
Fig. 8
D
E
E
E
E
cable tie
right side
cable tie left side