Replacement and repairs on the collimator 9 - 5
Siemens-Elema AB SPR8-215.898.01 Page 5 of 8 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
• Disconnect the cable to the connection terminal board "D" (Fig. 8).
• If there are cable ties, cut them (Fig. 8).
• Remove the collimator by loosening the four screws "E" (Fig. 8). Reserve the hardware.
• Install the new collimator. Center the rotary-flange so that the screws marked "E" are
centered in the grooves before tightening.
• Connect the wires to connection terminal board "D".