3 - 1
Siemens-Elema AB SPR8-215.898.01 Page 1 of 2 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Trouble-shooting chart 3
This chart shows the sequence beginning with the generator start, the exposure release
and ends with the developed X-ray film. Potential errors are mentioned in the function
box.
Additional hints that may be helpful during troubleshooting are listed under A-E in the sec-
tion "Troubleshooting instructions".
Power plug inserted
MOBILETT Plus,
MOBILETT Plus E,
MOBILETT Plus M
ON
Display illuminated
kV and mAs
buttons OK?
Collimator
lamp can be turned
ON/OFF
Turn plate
can be activated/
deactivated
Green
Ready light
indicator is ON
Voltage supply
Program run,
see section "P04 - Display, led
and buzzer test"
Connections
Button function
Defective lamp
Cables
Fuse
Lamp button
remote control (option)
Cables
Fuse
Turn plate buttons
Defective magnet
Charging time
Displayed message
not acknowledged
No or partly
No
No
No
No
Yes
Yes
Yes
Yes
Yes
A
B
C
D
E
Yes