HT and single tank error diagnosis 6 - 5
Siemens-Elema AB SPR8-215.898.01 Page 5 of 12 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Exposure values: 40 kV 5 mAs, HT 8 kV/div (200 mV/div),
I
inverter-single tank
100 A/div (only the end of the exposure is shown)
.
Fig. 2 HT and inverter current, inverter and single tank normal
Inverter
current
HT on
D1