Mechanical systems 10 - 17
Siemens-Elema AB SPR8-215.898.01 Page 17 of 28 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
• Switch on the light localizer and check the coincidence of the light and radiation fields
according to the specifications in the section on "Replacement of collimator".
• Reattach all covers on the single tank.
• Adjust the adjustment screw so that the single tank is balanced. If any problems occur,
refer to the section "Adjustment of arm system".
• Use the service program "P16 - Show the exposure counter" and record the values of the
exposure counter in the service protocol and in the unit documentation.
• Use the service program "P10 - Fast adaptation" to adjust the filament values to the new
single tank.
• Reattach the remaining covers.