Mechanical systems 10 - 21
Siemens-Elema AB SPR8-215.898.01 Page 21 of 28 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
• Set S10 to the OFF position.
• Reinstall all covers except single tanks covers "A" and "B".
• Make sure that the screw mountings between the fork and the arm are tightened
properly.
• Remove the rope between the handle and the foot while keeping the single tank on the
table.
• Tighten the adjustment screw until the single tank and arm are balanced. Refer to the
section "Adjustment of arm system".
• Switch on the light localizer and check the coincidence of the light and radiation fields
according to the specifications in the section "Replacement of collimator".
• Reattach the remaining covers.
Do not stretch the cable too tightly in the arm.
There must be enough slack so that the cable is not pulled by the
mechanical movements.