SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 4 of 22 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
5 - 4 Adjustments / Programs
Error examples:
Fig. 1
kV-act (20 kV = 1.0 V)
mA-act (2.5 mA = 1.0 V)
relevant time period
Tube current exceeds the upper limit
- preset filament current too high
13 mA
6 ms
14.5 mA
11.5 mA
3 ms
Tube current exceeds the lower limit
- preset filament current too low
13 mA
6 ms
14.5 mA
11.5 mA
3 ms
Fig. 1a
Fig. 1b