SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 8 of 22 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
5 - 8 Adjustments / Programs
Measurement 5
Switch on fluoroscopy.
Read and record the value displayed on the dose rate meter.
If the value measured exceeds 9R/min, the maximum tube current must be reduced. If
the value measured is less than 9R/min, the tube current does not need to be reduced.
Switch fluoroscopy off.
Calculate the maximum tube current.
Multiply the tube current to be programmed by 10, since entry on the service PC must be
made in 1/10mA. Round off the value to the nearest whole number.
Fig. 2
If the maximum tube current was already reduced, the mA value
displayed at 110kV should be recorded and used for the calcula-
tion. The maximum skin dose rate to be set should equal 9R/min.
tube current displayed at 110kV in mA * 9 R/min
= tube current to be programmed in mA
measured dose rate value in R/min
Example: Tube current displayed at 110kV: 5mA
Maximum skin dose rate: 9R/min
Measured dose rate value : 10.2 R/min.
5 (mA) * 9 (R/min)
10.2 (R/min)
= 4.41 (mA) Rounded to the nearest whole number: 44
The value to be programmed is 44.
Preparations 5
Attach the dose ionization chamber
at a distance of 30 cm from the I.I.
housing (Fig. 3).
Select one of the FL curves S2 / HC2
or IODINE according to the
programming.
(Push-key)
Briefly release fluoro.
Select kV stop mode and enter
110kV (5mA)
30 cm
Input
screen
Chamber
Holder
Sirephos
NOTE