Adjustments / Programs 5 - 9
Siemens AG Register 5 RXR2-130.061.02 Page 9 of 22 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Programming 5
Connect the service PC to the serial interface of board D1.
Select the Adjustment, Parameters menu.
In the Combo box "Parameter Groups", select "Limitation of skin dose rate (FL)" .
Click on "Get from unit" .
Enter the value to be programmed in "Value" .
Click on "Set Value".
Select "Put to unit" . Read the menu guide on the service PC.
Select "Save to file".
The value to be programmed has been transferred to the SIREMOBIL Compact and
temporarily stored to the disk.
Checking 5
After programming, check the maximum skin dose rate again.
You must exit the service program with "Logoff" and "Quit". Only
then will the modified parameters be stored on the floppy disk as
a new file.