Loading...
Replacing boards / replacing components 6 - 15
Siemens AG Register 5 RXR2-130.061.02 Page 15 of 18 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Perform the IQ quick test.
Memoskop C 6
Replace the Memoskop C.
The attachment screws are accessible on the underside of the storage device.
Set the Memoskop C to the correct video frequency and configure it for the number of
existing monitors.
Refer to Chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors".
Configure the following parameters in Technical Setup for Memoskop:
E Monitor split Configure horizontally. Configure vertically only upon cus-
tomer request.
F Time Configure Yes.
G Date / Time setting Set the current date and time.
H Hospital name Program the hospital name.
I Storing the dose If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise, configure No.
K Dose documentation If the area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Read and record the parameters for Memoskop C if you can
select User setup. After replacing the Memoskop, configure the
Memoskop C-E according to the parameters noted.
3 Number of monitors Configure 1 for SIREMOBIL Compact with 1 monitor.
Configure 2 for SIREMOBIL Compact with 2 monitors.
4 Hardcopy setup Check whether 5500 is programmed in "4 Max Hardcopy
time". If not, program this value.
5 Edge 1 coef set Check whether 3 is programmed, If not, program this value.
6 Edge 2 coef set Check whether 8 is programmed, If not, program this value.
7 Motion detection setup Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
10 Date / Time format Set the country-specific format.
12 Keyboard Set to the keyboard layout being used.
13 Language Set for language of the country.
15 Display of realtime clock Configure Yes. Select No only upon customer request.
16 Display of rotated images Configure Yes.
17 Monitor zoom Configure No.
18 Monitor split Configure Yes for SIREMOBIL Compact with 1 monitor.
Configure No for SIREMOBIL Compact with 2 monitors.
19 Monitor text Configure Yes.
20 Maximum image number Configure 100.