Adjustments / Programs 5 - 3
Siemens AG Register 5 RXR2-130.061.02 Page 3 of 22 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Checking the generator adjustment 5
With the system on, connect the service PC to the SIREMOBIL Compact.
Select "Test..." in the Diagnostic menu.
Select cassette simulation in "Function Groups".
Select cassette switch ON in "Available Functions".
The SIREMOBIL Compact is now in direct exposure mode.
Select 81 kV with the kV+ / kV- keys.
Select 8 mAs with the mAs+ / mAs- keys.
Connect oscilloscope channel 1 to D1.X46, ground at 0V. Set the sensitivity to 1V /Div.
Connect oscilloscope channel 2 to D1.X48, ground at 0 V. Set the sensitivity to 2.5V/ Div.
Set the time to 25 ms / Div.
Set the oscilloscope to storage mode.
Release an exposure with the hand switch and store the oscillogram with the exposure.
Evaluate the stored oscillograms.
The tube current must reach the tolerance range of 13 mA +/- 1.5 mA in the time of 3 ms
and may not leave this range for the next 6 ms.
(The tube current may not run above 14.5 mA within the first 6 ms).
Refer also to oscillograms in Fig. 1 / Fig. 1a and Fig. 1b.
Select "Cassette switch off" on the service PC in "Available Functions".
Direct exposure mode is cancelled.
If the unit exceeds the tolerance range, the generator adjustment must be repeated.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING