SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 6 of 18 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
6 - 6 Replacing boards / replacing components
• Remove the counterbalancing weight on the 7”/17cm I.I. (1/Fig. 4).
• Remove the plug from the control section on the I.I. mini voltage supply (1/Fig. 1/2).
• Disconnect the protective conductor connections and the grounding screw (5/Fig. 1).
• Remove plug D4.X1 of the Videomed DC (2/Fig. 2).
• Secure the end of the cable which you have wound up on the cable reel of the Videomed
DC with adhesive tape.
• Remove the cable between the Videomed DC and the Compact optics and cut the cable
ties securing it.
• Remove plug X45 of the cassette contact. (2/Fig. 2).
• Remove both jacks D4.X1 and X45 from the receptacle (2/Fig. 2).
Fig. 1 Fig. 2
Fig. 3 Fig. 4