SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 6 of 18 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
6 - 6 Replacing boards / replacing components
Remove the counterbalancing weight on the 7”/17cm I.I. (1/Fig. 4).
Remove the plug from the control section on the I.I. mini voltage supply (1/Fig. 1/2).
Disconnect the protective conductor connections and the grounding screw (5/Fig. 1).
Remove plug D4.X1 of the Videomed DC (2/Fig. 2).
Secure the end of the cable which you have wound up on the cable reel of the Videomed
DC with adhesive tape.
Remove the cable between the Videomed DC and the Compact optics and cut the cable
ties securing it.
Remove plug X45 of the cassette contact. (2/Fig. 2).
Remove both jacks D4.X1 and X45 from the receptacle (2/Fig. 2).