1.9 Maintenance interval
1.9.1 Maintenance interval, system
NOTE
The maintenance interval applies to all maintenance work, except
for the maintenance of the 3D reconstruction in combination with
a navigation system.
12 months
1.9.2 Maintenance interval, 3D reconstruction combined with navigation
system
NOTE
If the system is equipped with the 3D reconstruction option in
combination with a navigation system, perform the maintenance
work listed below in the specified maintenance interval.
NOTE
Take into account that the maintenance work listed below has to
be performed together with an engineer of the navigation system
manufacturer.
6 months, however after 150 patients (approx. 300 3D scans) at the latest.
SIM Navigation system (if present)
I.I. ring damage
Correction of the seating of the I.I. cap
SIE 3D reconstruction (if present)
Calibration of 3D reconstruction
Check of the limiting resolution
SIE 3D reconstruction with navigation system (if present)
Check the overall navigation accuracy