4.3 Functional inspection
PMF Video printer
Check the function of the video recorder.
Check the remote control (Print button).
PMF Video recorder
Check the function of the video recorder.
Check the function of the remote control (REC-Pause-Start / REC-Pause-Stop).
PMF Laser camera system "Sneaker Net"
Check the function of the laser-camera system “Sneaker Net”.
PMF DICOM bridge
Check the function of the DICOM bridge.
PMF DICOM Connect
Check the function of the DICOM Connect network connection.
PMF Codonics printer manual control panel
Check the function of the printer.
Check the function of the manual control panel.
Inspect the cable of the manual control panel for damage and breaks.
PMF 4 View
Check the plug-in connections for damage.
Check the function of the 4 View monitor extensions.
Check the function of the additional monitors and the additional keyboard (if present).
PMF Laser camera connection
Check the functioning of the laser camera connection.
Check the remote control (Print button).
PMF Image rotation
X
Check the image rotation (camera rotation).
Check the 0° position and the entire rotation range ( ± 220°) under fluoroscopy.
Check the function of the image rotation.
PMF Slot diaphragm
Select I.I. full format.
X
Open the slot diaphragm completely and briefly release fluoroscopy.
The slot diaphragm plates are no longer visible on the screen.
Select I.I. zoom format.
X
Release fluoroscopy briefly.
The slot diaphragm plates are no longer visible on the edges of the image.