Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C
07.05 CS PS 24
Maintenance, Operating Value / Functional Inspection 41
Page 41 of 46
Medical Solutions
4.3 Functional inspection
PMF Video printer
• Check the function of the video recorder.
• Check the remote control (Print button).
PMF Video recorder
• Check the function of the video recorder.
• Check the function of the remote control (REC-Pause-Start / REC-Pause-Stop).
PMF Laser camera system "Sneaker Net"
• Check the function of the laser-camera system “Sneaker Net”.
PMF DICOM bridge
• Check the function of the DICOM bridge.
PMF DICOM Connect
• Check the function of the DICOM Connect network connection.
PMF Codonics printer manual control panel
• Check the function of the printer.
• Check the function of the manual control panel.
• Inspect the cable of the manual control panel for damage and breaks.
PMF 4 View
• Check the plug-in connections for damage.
• Check the function of the 4 View monitor extensions.
• Check the function of the additional monitors and the additional keyboard (if present).
PMF Laser camera connection
• Check the functioning of the laser camera connection.
• Check the remote control (Print button).
PMF Image rotation
X
• Check the image rotation (camera rotation).
Check the 0° position and the entire rotation range ( ± 220°) under fluoroscopy.
• Check the function of the image rotation.
PMF Slot diaphragm
• Select I.I. full format.
X
• Open the slot diaphragm completely and briefly release fluoroscopy.
The slot diaphragm plates are no longer visible on the screen.
• Select I.I. zoom format.
X
• Release fluoroscopy briefly.
The slot diaphragm plates are no longer visible on the edges of the image.