Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C
07.05 CS PS 24
Safety Inspection 29
Page 29 of 46
Medical Solutions
3-
3 Safety Inspection
3.1 Mechanical safety
SIM Cover panels
• Remove the covers from the SIREMOBIL stand and monitor trolley.
• Check cover panels for damage.
SIM Cable deflectors
• Inspect the cable deflectors on the stand and on the monitor trolley.
SIM Foot brake
• Test the function of the foot brakes on the SIREMOBIL stand and monitor trolley on a
level floor.
SIM Brakes
• Use the spring balance to test whether the orbital, angulation, swivel and horizontal
movement brakes apply the prescribed force. (Refer to service instructions SIREMO-
BIL ISO-C)
SIM C-arm
• Check the C-arm for damage to the carbon fiber structures (visual inspection). If there is
damage, proceed as described in the warning in chapter 1.
• Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
• Orbital movement: Once the orbital brake is released, the C-arm can be moved only in a
limited manner. Check whether the orbital movement is smooth.
SIM Wheels
• Move the SIREMOBIL in a straight line on a level floor.
• Evaluate the straight and quiet movement of the SIREMOBIL.
• Replace any defective wheels or castors.
• Check the steering function.
SIM Lifting column
• Switch the system on
• Electrically raise and lower the lifting column,
- while listening for noises and checking for play in the bearings.
- The lifting column movement must switch off automatically when the end positions
are reached.
• Switch off the system.
• Lubricate the lifting column.
- Lubricate the bearing surfaces of the lifting column guide and the spindle.
SIM Emergency Stop switch
• Press the Emergency Stop switch. The lifting column may not be raised or lowered.
• Press the Emergency Stop switch again. The lifting column can be moved again.