EasyManua.ls Logo

Siemens SIREMOBIL Iso-C

Siemens SIREMOBIL Iso-C
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C
07.05 CS PS 24
Safety Inspection 33
Page 33 of 46
Medical Solutions
Check the iris collimator and correct it if necessary.
X - Select the full I.I. format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
- Select a zoom format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
SIE Area-dose product measuring device
Check the area-dose product measuring device
X
(refer to the service instructions for SIREMOBIL ISO-C).
SIE Codonics printer cable break
Inspect all connection lines between the printer and the Framegrabber for damage and
breaks. Inspect the cable routing of the energy chain for damage and proper functional-
ity (ensure that none of the cables are pinched).
3.2.1 Monitor(s)/displays
SIE Monitor(s)/displays
Check the SIMOMED N monitors as prescribed in the checklist.
Check the luminance of the standard monitors with 100/120 Hz.
Check the luminance of the TFT displays. Check the ambient light sensor.

Table of Contents

Other manuals for Siemens SIREMOBIL Iso-C

Related product manuals