Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C
07.05 CS PS 24
Safety Inspection 33
Page 33 of 46
Medical Solutions
• Check the iris collimator and correct it if necessary.
X - Select the full I.I. format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
- Select a zoom format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
SIE Area-dose product measuring device
• Check the area-dose product measuring device
X
• (refer to the service instructions for SIREMOBIL ISO-C).
SIE Codonics printer cable break
• Inspect all connection lines between the printer and the Framegrabber for damage and
breaks. Inspect the cable routing of the energy chain for damage and proper functional-
ity (ensure that none of the cables are pinched).
3.2.1 Monitor(s)/displays
SIE Monitor(s)/displays
• Check the SIMOMED N monitors as prescribed in the checklist.
• Check the luminance of the standard monitors with 100/120 Hz.
• Check the luminance of the TFT displays. Check the ambient light sensor.