Siemens AG SPR2-230.101.02.06.02 SIREMOBIL Iso-C
07.05 CS PS 24
Safety Inspection 37
Page 37 of 46
Medical Solutions
3.6 Measurements
SIE Protective conductor test (performed under "Final Work Steps")
Test the protective conductor with the protective conductor wire tester according to
ARTD-002.731.17... .
The protective conductor resistance must not exceed 0.2 Ohms.
NOTE
The protective conductor test is part of the safety inspection. How-
ever, it cannot be performed until all maintenance work has been
completed. See (Final Work Steps / p. 45).
SIE Leakage current test (performed under "Final Work Steps")
Perform the leakage current test with the system closed according to
ARTD-002.731.17. Document and evaluate the determined values. Document the
measuring procedure. The leakage current must not exceed 2.5 mA.
NOTE
The leakage current test is part of the safety inspection. However,
it is recommended to perform it only at the end of maintenance,
with the system closed. See (Final Work Steps / p. 45).