30 Safety Inspection
SIREMOBIL Iso-C SPR2-230.101.02.06.02 Siemens AG
07.05 CS PS 24
Page 30 of 46
Medical Solutions
3.1.1 I.I. laser light localizer
SIM Mechanical system
• Inspect the I.I. laser light localizer for mechanical damage.
• Mount the I.I. laser light localizer on the I.I. and ensure proper locking and seating.
When doing this, pay special attention to the tension band and its closure.
• Check the I.I. ring for damage.
SIM Function
• Perform maintenance on the I.I. laser light localizer according to Installation and Set-
ting Instructions RXR2-130.033.03...
3.1.2 Laser light localizer
SIM Mechanical system
• Inspect the I.I. cap of the laser light localizer for mechanical damage.
• Install the I.I. cap of the laser light localizer on the I.I. and ensure proper locking and
seating.
• Check the I.I. ring for damage.
SIM Function
• Perform the maintenance steps specified according to the adjustment instructions
SPR2-230.032.01..
3.1.3 DICOM bridge
SIM Mounting
• Check the mounting of the DICOM bridge in the monitor trolley.
3.1.4 Codonics printer
SIM Mounting
• Inspect the screws used to fasten guide rail to the mounting bracket of the Framegrab-
ber.
• Ensure that the printer is properly fastened to the guide rail. Make sure there are no
missing attachment screws.
• Ensure that the Framegrabber is properly fastened to the mounting bracket. Make sure
there are no missing attachment screws.
SIM Guide rails
• Inspect the guide rails for proper function and smooth operation.
• The guide rails must not be damaged.
• The guide rails must lock properly into place when fully pushed into position.
• If necessary, use a cleaning brush to remove dirt from the rails.