Isolate the sensor from cathodic protected pipes using insulated bolts.
Use bypass cable between the mating flanges.
Note
Remote mounted sensor versions
If the above is not acceptable, remote mounted sensors can alternatively be connected as
follows:
Connect coil current cable shield at sensor end via a 1.5 µF condensator
Make sure that electrode cable shield is not connected at both ends
Installing/Mounting
4.8 Potential equalization
SITRANS F M MAG 1100 sensor
32 Operating Instructions, 01/2018, A5E03433301-AC