8.2 Rated operating conditions
Table 8-2 Ambient temperatures
1
Version MAG 1100 MAG 1100 HT MAG 1100 F
Standard Sensor -40 to +100 ºC (-40 to +212 °F) -40 to +100 ºC (-40 to +212 °F) -40 to +100 ºC (-40 to +212 °F)
Ex Sensor -20 to +60 ºC (-4 to +140 °F) -20 to +60 ºC (-4 to +140 °F) -20 to +60 ºC (-4 to +140 °F)
With compact
transmitter
MAG
5000/6000
-20 to +60 ºC (-4 to +140 °F) -20 to +60 ºC (-4 to +140 °F)
MAG 6000 I -20 to +60 ºC (-4 to +140 °F) -20 to +60 ºC (-4 to +140 °F)
MAG 6000 I
Ex de
-10 to +60 ºC (+14 to +140 °F) -20 to +60 ºC (-4 to +140 °F)
1
Conditions also dependent on liner characteristiques
Table 8-3 Media temperature
Version MAG 1100 MAG 1100 HT MAG 1100 F
Ceramic -20 to +150 °C (-4 to +302 °F)
MAG 1100 Ex
-20 to +150 °C (-4 to +302 °F)
-20 to +200 °C (-4 to +392 °F)
MAG 1100 Ex
-20 to +180 °C (-4 to +356 °F)
-20 to +150 °C (-4 to +302 °F)
MAG 1100 Ex
-20 to +150 °C (-4 to +302 °F)
PFA
-30 to +130 °C (-22 to +266 ºF)
Suitable for steam sterilization at
150 °C (302 °F)
-30 to +130 °C (-22 to +266 ºF)
Suitable for steam sterilization at
150 °C (302 °F)
8.3 Temperature shock
Table 8-4 Temperature shock
Version MAG 1100 MAG 1100 HT MAG 1100 F
Ceramic
Duration ≤ 1 min,
followed by 10
min rest
DN 2 and 3 (1/12 " and 3/8"):
No limitations
DN 6, 10, 15 and 25:
Max ΔT ≤ 80 °C/min
(1/4", 3/8", ½" and 1":
Max ΔT ≤ 144 °F/min)
DN 40, 50 and 65:
Max ΔT ≤ 70 °C/min
(1½", 2", and 2½":
Max ΔT ≤ 126 °F/min)
DN 80 and 100:
Max ΔT ≤ 60 °C/min
(3" and 4": Max ΔT ≤ 108 °F/min)
DN 15 and 25:
Max ΔT ≤ 80 °C/min
(½" and 1": Max ΔT ≤ 144 °F/min)
DN 40 and 50:
Max ΔT ≤ 70 °C/min
(1½" and 2":
Max ΔT ≤ 126 °F/min)
DN 80 and 100:
Max ΔT ≤ 60 °C/min
(3" and 4": Max ΔT ≤ 108 °F/min)
DN 10, 15 and 25:
Max. ΔT ≤ 80 °C/min
(3/8”, ½” and 1”:
Max. ΔT ≤ 144 °F/min)
DN 40, 50 and 65:
Max. ΔT ≤ 70 °C/min
(1½”, 2” and 2½”:
Max. ΔT ≤ 126 °F/min)
DN 80 and 100:
Max. ΔT ≤ 60 °C/min
(3”, 4”: Max. ΔT ≤ 108 °F/min)
PFA Max. ±100 °C (210 °F)
momentarily
Max. ± 100 °C (212 °F)
momentarily
Technical data
8.3 Temperature shock
SITRANS F M MAG 1100 sensor
50 Operating Instructions, 01/2018, A5E03433301-AC