Table 8-5 Operating pressure
Version MAG 1100 MAG 1100 HT MAG 1100 F
Ceramic
● DN 2 to 65: 40 bar
(1/12" to 2½":: 580 psi)
● DN 80: 37.5 bar
(3": 540 psi)
● DN 100: 30 bar
(4": 435 psi)
Vacuum 1 x 10
-6
bar
abs
(1.5 x 10
-5
psi
abs
)
● DN 2 to 65: 40 bar
(1/12" to 2½": 580 psi)
● DN 80: 37.5 bar
(3": 540 psi)
● DN 100: 30 bar
(4": 435 psi)
Vacuum 1 x 10
-6
bar
abs
(1.5 x 10
-5
psi
abs
)
● DN 10 … 65: 40 bar
(3/8" to 2½":: 580 psi)
● DN 80: 25 bar
(3": 363 psi)
● DN 100: 25 bar
(4": 363 psi)
Vacuum 1 x 10
-6
bar
abs
(1.5 x 10
-5
psi
abs
)
PFA 20 bar to (290 psi)
Vacuum 0.02 bar
abs
(0.3 psi
abs
)
DN 80 to 100: CO
2
pressure max.
7 bar (101.5 psi)
20 bar to (290 psi)
Vacuum 0.02 bar
abs
(0.3 psi
abs
)
DN 80 to 100: CO
2
pressure max.
7 bar (101.5 psi)
Table 8-6 Mechanical load/Vibration
Version MAG 1100 MAG 1100 HT MAG 1100 F
● 18 to 1000 Hz random in x, y,
z, directions for 2 hours
according to EN 60068-2-36
● Sensor: 3.17 grms
● Sensor with compact MAG
5000/ 6000 mounted
transmitter: 3.17 grms
● Sensor with compact MAG
6000 I/ 6000 I Ex de mounted
transmitter: 1.14 grms
● For compact installation with
the MAG 6000 I, transmitter
to be supported to avoid
tension on sensor part
● 18 to 1000 Hz random in x, y,
z, directions for 2 hours
according to EN 60068-2-36
● Sensor: 3.17 grms
● 18 to 1000 Hz random in x, y,
z, directions for 2 hours
according to EN 60068-2-36
● Sensor: 3.17 grms
● Sensor with compact MAG
5000/ 6000 mounted
transmitter: 3.17 grms
● Sensor with compact MAG
6000 I/ 6000 I Ex de mounted
transmitter: 1.14 grms
● For compact installation with
the MAG 6000 I, transmitter
to be supported to avoid
tension on sensor part
Table 8-7 Mechanical load/Vibration
Version MAG 1100 MAG 1100 HT MAG 1100 F
EMC IP67 to EN 60529 (NEMA 4X), 1
mH
2
O for 30 min 2004/108/EC
IP67 to EN 60529 (NEMA 4X), 1
mH
2
O for 30 min 2004/108/EC
IP67 to EN 60529 (NEMA 4X), 1
mH2O for 30 min 2004/108/EC
Technical data
8.3 Temperature shock
SITRANS F M MAG 1100 sensor
Operating Instructions, 01/2018, A5E03433301-AC 51