1. Unscrew and remove terminal box lid.
2. Mount the two terminal blocks as shown and insert electrode cable plug (terminals 82,
0 and 83) and coil cable plug (terminals 85 and 86).
Terminals electrode cable: 82, 0 and 83
Terminals special electrode cable: 84, 83, 0, 82 and 81
Terminals coil cable: 85; 86
Note
Special electrode cable
Special electrode cable must be connected to terminals 84, 83, 0, 82 and 81.
Figure 5-1 Terminal box without blocks
3. Remove SENSORPROM
®
unit from sensor and mount it on connection plate in transmitter,
see relevant transmitter operating instructions.
4. Fit the ½" NPT or M20 cable glands for supply and output cables.
Connecting
5.2 Remote installation
SITRANS F M MAG 1100 sensor
36 Operating Instructions, 01/2018, A5E03433301-AC