Troubleshooting
9.4 Application information guide
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
71
Full scale frequency (10 000 Hz) Menu 4.2.5
Volt UM (magnetic current) (23 100 - 30 500) Menu 6.5.1
Volt Uref (magnetic ref. voltage) Menu 6.5.1
Volt Usig Menu 6.5.2
Volt Uel1
and DC voltage on electrodes measured on terminals 22 to
23 from sensor
Menu 6.5.3 Menu Terminal
Volt Uel1
and DC voltage on electrodes measured on terminals 22 to
24 from sensor
Menu 6.5.4 Menu Terminal
Coil value (terminal 5 and 6 on sensor)
Isolation value between terminals 5/6 to 7 (min 10 MOhm)
Ref-coil value between terminals 55 and 66 [Ohm]
Sample frequency (approx. 2 x magnetic frequency) Menu 6.5.6
Service info (coded service information) Menu 6.5.8
Cal. factor (1.0) Menu 6.7.1
CFH value (are sensor and transmitter data the same?) Menu 6.7.2
CFR value (are sensor and transmitter data the same?) Menu 6.7.2
ZPH value (are sensor and transmitter data the same?) Menu 6.7.3
Excitation frequency (bipolar PP) Menu 6.7.4
Grounding sensor (correct?) See system
information