Connecting
6
6.1 Basic safety notes
WARNING
Incorrect conduit system
Risk of explosion in hazardous areas as result of open cable inlet or incorrect conduit system.
• In the case of a conduit system, mount a spark barrier at a dened distance from the device
input. Observe national regulations and the requirements stated in the relevant approvals.
WARNING
Hazardous contact voltage
Risk of electric shock in case of incorrect connection.
• For the electrical connection specications, refer to the information in Technical
specications (Page45).
• At the mounting location of the device observe the applicable directives and laws for
installation of electrical power installations with rated voltages below 1000V.
WARNING
Connecting or disconnecting in explosive environments
Connecting or disconnecting a powered device in explosive environments can lead to an
explosion.
• Connect and disconnect in non-explosive environments.
- or-
• Remove power to the device before connecting or disconnecting in explosive atmosphere.
WARNING
Connecting or disconnecting device in energized state
Risk of explosion in hazardous areas.
• Connect or disconnect devices in hazardous areas only in a de-energized state.
• Install a suitable switch-o device.
Exceptions:
• Devices having the type of protection "Intrinsic safety Ex i" may also be connected in
energized state in hazardous areas.
SITRANS Vericator
Operating Instructions, 03/2024, A5E51481698-AA 23