Reference
Battery station
Power supply cable
AC power cable
Figure6-2 Battery station with cables
Charging the battery
Note
Charge battery prior to 󹪝rst use
On delivery the battery level will be low. It is recommended to fully charge the battery prior to
󹪝rst use.
Note
Ambient temperature
The ambient temperature during charging the battery has to remain in the speci󹪝ed range, as
described in Operating conditions (Page45).
1. Take or remove the battery. (Page27)
2. Plug the battery into battery station.
3. Connect the power supply cable to battery station.
4. Connect the AC power cable to power supply.
5. Plug the AC power cable into an AC power outlet.
LED Indicator
One time Red/Orange/Green Selftest: Charger is ready for use.
Red/Green blinking Battery recognition and initialization.
Orange light The inserted battery is of the correct type and is currently being charged.
Green light The battery is charged and can be removed for use.
Red blinking The battery is too hot or too cold to be charged without damage. If the battery is too cold
it will be charged as soon as it has warmed up su󹪟ciently. If the battery is too hot it
should be removed to cool down.
Red light The battery is damaged or it is a conventional battery which cannot be recharged.
Connecting
6.5Charging the battery