105
Shoulder
Shoulder
kV 130
Effective mAs 85
Slice collimation 2.0 mm
Slice width 3.0 mm
Feed/Rotation 12.0 mm
Rotation time 1.5 sec.
Kernel B60s
Increment 2.0 mm
Image order cr-ca
CTDI
Vol
9.7 mGy
In case of 3D study, mAs value can be reduced by 50%.
Use raw data to review a target region if necessary.
For image reconstruction of soft tissue use kernel B31s
and a slice width of 5.0 mm.