General
Kernels
There are 5 different types of kernels: “H“ stands for
Head, “B“ stands for Body, “U“ stands for High Reso-
lution, “C“ stands for ChildHead, and “S“ stands for
Special Application, e. g., Osteo CT.
The image sharpness is defined by the numbers
the higher the number, the sharper the image; the
lower the number, the smoother the image.
A set of 32 kernels is supplied with the Emotion 6 Soft-
ware, consisting of 13 body kernels B10s, B20s, B30s,
B31s, B35s, B40s, B41s, B46s, B50s, B60s, B70s, B80s,
B90s, 12 head kernels H10s, H20s, H21s, H30s, H31s,
H40s, H41s, H50s, H60s, H70s, H80s, H90s, 3 child
head kernels C20s, C30s, C60s, 4 special kernels S30s,
S80s, S90s, U90s.
Note: Do not use different kernels for other body parts
other than what they are designed for.
For further information regarding the kernels, please
refer to the “Hints in General” of the corresponding
body region.
Image Filter
There are 4 different filters available:
PFO: To reduce beam-hardening artifacts in head
images, particularly in the base of the skull, use the
Posterior Fossa Optimization (PFO) filter. It is highly
recommended to use the PFO filter for all soft tissue
reconstruction.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM Emotion 6 and is the answer not in the manual?

Siemens SOMATOM Emotion 6 Specifications

General IconGeneral
BrandSiemens
ModelSOMATOM Emotion 6
CategoryMedical Equipment
LanguageEnglish

Related product manuals