176
Spine
SpineRoutineSeq
Indications:
Sequence mode for spine studies, e. g. prolapse,
degenerative changes, trauma, tumors etc.
This protocol contains three ranges:
L3-L4, L4-L5, L5-S1.
The scan length of each range is 2.8 cm
Three different typical gantry tilts are predefined:
for L3-L4: 0°, for L4-L5: +5° and for L5-S1: +15°
SpineRoutineSeq
L3-4 L4-5 L5-S1
kV 130
mAs 280 280 320
Slice collimation 1.0 mm
Slice width 2.0 mm
Feed/Scan 6.0 mm
Rotation time 1.5 sec.
Kernel B31s
Cycle time 3.0 sec.
Image order cr-ca
CTDI
Vol
36.9 mGy 36.9 mGy 42.2 mGy